Siemens Healthcare will be informing customers about potential dot- or line-shaped artifacts caused by the presence of lubricating grease within the headband of the over-ear headphones listed below when used during head examinations with the Magnetom MRI…
RecallHealth product recall | 2023-09-11
The update provides a network isolator for the lan connection, which must be installed in the Mira Max vf10 systems. The isolator is a measure of operator protection.
Recall start date: May 29, 2023
RecallHealth product recall | 2023-06-08
The new software version VC20G provides latest cybersecurity updates and improvements in the following categories:
- Improves overall system stability
- Improves dicom connectivity
- Improves biopsy workflow
Specifically, the new software version…
RecallHealth product recall | 2023-05-11
The service pack (full version and delta package) NX VA20A-SP06 (VA20A-04T4) contains improvements concerning performance and stability.
- image reconstruction failure solved
- Petra sequence image text corrected
- new LW for RCCS_2G to support new…
RecallHealth product recall | 2023-04-13
Results of post market surveillance revealed that under certain circumstances the room configuration parameters of the system are set to default values. If these values are larger than the actual room dimensions, a collision with the ceiling or wall…
RecallHealth product recall | 2023-02-24
Siemens Healthcare is information customers of a potential issue with the Artis system in combination with a Siemens Healthineers table or a Trumpf/Maquet table. There are 3 potential issues depending on the combination of the Artis Pheno and type of…
RecallHealth product recall | 2023-02-10
In the Syngo.ct va40 and Syngo.ct va50 myneedle guide application, the order of the orientation labels "head", "center", and "feet" in the axial thick slice segments are erroneously swapped in the following two cases:
• if the recon parameter "image…
RecallHealth product recall | 2023-01-12
In rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. The full system functionality is not available. Only the backup mode is available, as mentioned in the operator manual. The Syngo start-…
RecallHealth product recall | 2023-01-03
During transversal table movement, it could happen that the rolling bearing which is part of the linear guide rail and enables the transversal table movement may get dislocated. The rolling bearing may fall out of the linear guide rail and thus the…
RecallHealth product recall | 2023-01-03
Sporadically, the distance measurement values displayed in MRI images on third-party postprocessing systems might be wrong if:
- enhanced MRI images with rectangular FOV (field of view) are created at the MR scanner, and
- in Syngo.via or MR view&…
RecallHealth product recall | 2023-01-03
Results of post market surveillance revealed that under certain circumstances the room configuration parameters of the system are set to default values. If these values are larger than the actual room dimensions, a collision with the ceiling or wall…
RecallHealth product recall | 2022-12-30
Siemens Healthcare is informing customers with a preventive measure to exclude any potential risk of patient injury caused the footrest detaching from the patient table during an examination.
Recall start date: December 15, 2022
RecallHealth product recall | 2022-12-23
Issue 1: "password store corrupted" error message during system bootup in rare cases (e.g., if an unplanned and unguided shutdown is triggered via the power button) it is possible that during a later syngo start-up an error message occurs instead of the…
RecallHealth product recall | 2022-12-02
The Artis Pheno System is provided with a dedicated error management mechanism. In rare cases of specific X-ray tube failures, the system is intended to automatically switch over to "bypass fluoroscopy" mode. However, in case such a failure occurs while…
RecallHealth product recall | 2022-09-28
The Artis system is provided with a dedicated error detection mechanism. In rare cases of failure of this error detection mechanism it will not be possible to release x-ray anymore until system shutdown.
Recall start date: August 12, 2022
RecallHealth product recall | 2022-09-20