Siemens Biograph mMR System
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Biograph mMR System |
151022 |
10433372 |
Issue
Siemens Healthcare GmbH became aware of this issue as part of a customer complaint investigation. It has been identified that this issue can occur when certain parameters (Echo spacings 0.50 MS; 0.51 MS, and 0.52 MS) are used in EPI (Echo-Planar Imaging), sequences will excite high internal stress inside the gradient coil assembly. This may lead to cracks and damage to the gradient coil.
Recall start date: March 15, 2024
Additional information
Details
Siemens Healthcare GmbH
Henkestr. 127, Erlangen, Germany, 91052
Get notified
Receive emails about new and updated recall and safety alerts.