Help us improve our website

Health product recall

Magnetom System

Last updated

Summary

Product
Magnetom System
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalog number

Magnetom Altea

Not applicable

11344915

Magnetom Sola Fit

Not applicable

11410482

Issue

The service pack (full version and delta package) NX VA20A-SP06 (VA20A-04T4) contains improvements concerning performance and stability.
- image reconstruction failure solved
- Petra sequence image text corrected
- new LW for RCCS_2G to support new flash HW components, new LW introduced for Sep 025 VA03A

To handle short power loss, new LW to support B2B magnet for Vida
- additional SW-freeze and reboot use cases are improved
- upgradeability directly from XA20A to XA50A
- data security hotfixes

Recall start date: April 1st, 2023

Additional information

Details
Original published date: 2023-04-13
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Siemens Healthcare GmbH

Henkestr. 127, Erlangen, Germany, 91052

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-73499