Magnetom Systems
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Magnetom Lumina | 196408 196406 196427 | 11344916 |
Magnetom Skyra | 45186 45281 45446 45166 45087 145249 45768 145034 45934 45961 45370 145335 45447 | 10432915 |
Magnetom Sola Fit | 202567 202589 202562 202570 | 11410482 |
Magnetom Vida | More than 10 numbers, contact manufacturer. | 11060815 |
Magnetom Altea | 189602 189915 189956 189700 190027 | 11344915 |
Magnetom Sola | More than 10 numbers, contact manufacturer. | 11291455 |
Magnetom Aera | More than 10 numbers, contact manufacturer. | 10432914 |
Magnetom Prisma | 166026 166139 166247 66071 166208 66108 | 10849582 |
Magnetom Avanto Fit | 211019 | 11516216 |
Issue
Sporadically, the distance measurement values displayed in MRI images on third-party postprocessing systems might be wrong if:
- enhanced MRI images with rectangular FOV (field of view) are created at the MR scanner, and
- in Syngo.via or MR view&go, the user rotates the images to display them in any orientation and saves them, and
- afterwards, the user exports the images to another Dicom node using MR export rules (interoperability mode) the distance measurement values on newly created series might be displayed incorrectly on systems that receive these new images.
Recall start date: December 21, 2022