Artis Icono and Pheno Systems
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Artis Icono Biplane |
180415 |
11327600 |
Artis Pheno |
164101 |
10849000 |
Artis Icono Floor |
170369 |
11327700 |
Issue
During transversal table movement, it could happen that the rolling bearing which is part of the linear guide rail and enables the transversal table movement may get dislocated. The rolling bearing may fall out of the linear guide rail and thus the mechanical connection between table top and table base may get lost.
Recall start date: December 19, 2022
Additional information
Details
Siemens Healthcare Gmbh
Henkestr. 127, Erlangen, Germany, 91052
Get notified
Receive emails about new and updated recall and safety alerts.