Somatom X.Cite
Brand(s)
Summary
Contact the manufacturer if you require additional information
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Somatom X.Cite |
123296 |
11330001 |
Issue
In the Syngo.ct va40 and Syngo.ct va50 myneedle guide application, the order of the orientation labels "head", "center", and "feet" in the axial thick slice segments are erroneously swapped in the following two cases:
• if the recon parameter "image order" is changed by the user from the default setting "head to feet" to "feet to head", or
• if the recon parameter "mirroring" is changed by the user from the default setting "none" to any other value, this influences the parameter "image order", which is switched from the default setting "head to feet" to "feet to head" so that above-mentioned condition applies.
Recall start date: January 5, 2022
Additional information
Details
Siemens Healthcare Gmbh
Henkestr. 127, Erlangen, Germany, 91052
Get notified
Receive emails about new and updated recall and safety alerts.