Health product recall

Somatom X.Cite

Last updated

Summary

Product
Somatom X.Cite
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Somatom X.Cite

123296

11330001

Issue

In the Syngo.ct va40 and Syngo.ct va50 myneedle guide application, the order of the orientation labels "head", "center", and "feet" in the axial thick slice segments are erroneously swapped in the following two cases:

• if the recon parameter "image order" is changed by the user from the default setting "head to feet" to "feet to head", or

• if the recon parameter "mirroring" is changed by the user from the default setting "none" to any other value, this influences the parameter "image order", which is switched from the default setting "head to feet" to "feet to head" so that above-mentioned condition applies.

Recall start date: January 5, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Siemens Healthcare Gmbh

Henkestr. 127, Erlangen, Germany, 91052

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72559