Health product recall

Artis Icono Ceiling

Last updated

Summary

Product
Artis Icono Ceiling
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Artis Icono Ceiling

190027
190036

11328100

Issue

Artis Icono Ceiling systems are equipped with an automatic collision detection at locations where collisions with users or equipment might occur. If the issue occurs, the collision sensor around the x-ray tube housing shows the following unintended behavior: when the Artis Icono Ceiling system was switched off for a longer time (e.g. overnight) and switched on, this collision sensor is falsely signaling non-existing collisions for up to half an hour.

Recall start date: October 19, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Siemens Healthcare Gmbh
Henkestr. 127, Erlangen, Germany, 91052
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74549

Get notified

Receive emails about new and updated recall and safety alerts.