We have become aware of a malfunction with the Hamilton-C2/C3/C1/T1 ventilators when used long term in neonatal patient group. If an affected device is used accumulatively for 91 days in neonatal patient group without a restart of the device, the device…
RecallHealth product recall | 2023-09-26
Hamilton Medical Ag is recalling certain of the ventilators' Hamilton-C1/T1/mr1 control boards after receiving reports that the device may interrupt ventilation due to a faulty capacitor.
Recall start date: August 1, 2023
RecallHealth product recall | 2023-08-21
Two potential Hamilton-C6 Ventilator malfunctions detected following an inspection by Hamilton Medical Ag.:
Malfunction 1: Safety ventilation can be triggered by switching to an adaptive mode.
Malfunction 2: The display can be restarted by…
RecallHealth product recall | 2023-03-23
Status indicator board dislodging with potential to cause ventilator to shut down
Recall start date: May 27, 2022
RecallHealth product recall | 2022-06-07