Hamilton C6 Intensive Care Ventilator
Brand(s)
Dernière mise à jour
Summary
Produit
Hamilton C6 Intensive Care Ventilator
Problème
Medical devices - Performance issue
Ce qu’il faut faire
Instructions provided by manufacturer to mitigate. Software update to be provided later.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Hamilton C6 Intensive Care Ventilator |
Contact manufacturer. |
160021 |
Problème
Two potential Hamilton-C6 Ventilator malfunctions detected following an inspection by Hamilton Medical Ag.:
Malfunction 1: Safety ventilation can be triggered by switching to an adaptive mode.
Malfunction 2: The display can be restarted by changing the target shift setting (only in intellivent-asv mode).
Recall start date: March 1, 2023
Additional information
Details
Original published date: 2023-03-23
Type d’avis ou de rappel
Health product recall
Catégorie
Health product - Medical device - Anaesthesiology
Entreprises
Hamilton Medical Ag
Via Crusch 8, Bonaduz, Switzerland, 7402
Published by
Health Canada
Audience
Healthcare
Classe de rappel
Type II
Identification number
RA-73374