Hamilton C6 Intensive Care Ventilator
Issue
Medical devices - Performance
What to do
Instructions provided by manufacturer to mitigate. Software update to be provided later.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Hamilton C6 Intensive Care Ventilator |
Contact manufacturer. |
160021 |
Issue
Two potential Hamilton-C6 Ventilator malfunctions detected following an inspection by Hamilton Medical Ag.:
Malfunction 1: Safety ventilation can be triggered by switching to an adaptive mode.
Malfunction 2: The display can be restarted by changing the target shift setting (only in intellivent-asv mode).
Recall start date: March 1, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Hamilton Medical Ag
Via Crusch 8, Bonaduz, Switzerland, 7402
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73374