Health product recall

Hamilton C6 Intensive Care Ventilator

Dernière mise à jour

Summary

Produit
Hamilton C6 Intensive Care Ventilator
Problème
Medical devices - Performance issue
Ce qu’il faut faire

Instructions provided by manufacturer to mitigate. Software update to be provided later. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Hamilton C6 Intensive Care Ventilator

Contact manufacturer.

160021

Problème

Two potential Hamilton-C6 Ventilator malfunctions detected following an inspection by Hamilton Medical Ag.: 

Malfunction 1: Safety ventilation can be triggered by switching to an adaptive mode.
Malfunction 2: The display can be restarted by changing the target shift setting (only in intellivent-asv mode).

Recall start date: March 1, 2023

Additional information

Details
Original published date: 2023-03-23
Type d’avis ou de rappel
Health product recall
Catégorie
Health product - Medical device - Anaesthesiology
Entreprises

Hamilton Medical Ag

Via Crusch 8, Bonaduz, Switzerland, 7402

Published by
Health Canada
Audience
Healthcare
Classe de rappel
Type II
Identification number
RA-73374