Hamilton C6 Intensive Care Ventilator
Brand(s)
Last updated
Summary
Product
Hamilton C6 Intensive Care Ventilator
Issue
Medical devices - Performance
What to do
Inform the distributor immediately if you detect the loosening of the status indicator board.
Audience
Healthcare
Affected products
Affected products |
Lot or Serial number |
Model or Catalog number |
|---|---|---|
Hamilton C6 Intensive Care Ventilator |
More than 10 numbers, contact manufacturer. |
160021 |
Issue
|
Status indicator board dislodging with potential to cause ventilator to shut down Recall start date: May 27, 2022 |
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Hamilton Medical Ag
Via Crusch 8, Bonaduz, Switzerland, 7402
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64237
Get notified