In2Flow Nasal Cannulas
Brand(s)
Summary
Customers provided instructions on how to respond when this becomes a problem.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
In2Flow Nasal Cannulas |
1900059952 |
10076606 |
Issue
Failure description: several complaints were received where the plug (the one at the opposite side of the tubing) became detached from the In2Flow nasal cannula, causing leakage due to unintentional disconnection of that part. This may result in degraded high flow support and oxygen delivery, which may lead to patient O2 desaturation.
Recall Start Date: August 31, 2023
Additional information
Details
Hamilton Medical AG
Via Crusch 8, Bonaduz, Switzerland, 7402
Get notified
Receive emails about new and updated recall and safety alerts.