Health product recall

Hamilton C2 and C3 Intensive Care Ventilators

Last updated

Summary

Product
Hamilton C2 and C3 Intensive Care Ventilators
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Hamilton-C2 Intensive Care Ventilator

Contact the manufacturer.

160002

Hamilton-C3

Contact the manufacturer.

160005

Issue

This field safety notice (FSN) provides an updated information on Hamilton medical ventilators Hamilton-C2/C3: Hamilton Medical AG became aware of a malfunction with our Hamilton-C2/C3 ventilators when used long term in neonatal patients.  If an affected device Hamilton-C2 or Hamilton-C3 is used accumulatively for 91 days in neonatal patient without a restart of the device, the device will switch to "ambient state" and therefore stop ventilation.  During the "ambient state" the device is alarming audibly until it is switched off by the user.  Hence, it is mandatory to apply a periodic preventive restart of affected devices.

Devices and software version affected:

  • All Hamilton-C2 with sw version ≤2.2.5 and neonatal option installed
  • All Hamilton-C3 with sw version ≤2.0.9 and neonatal option installed

Recall start  date: January 29, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Hamilton Medical AG

Via Crusch 8, Bonaduz, Switzerland, 7402

Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-76965