Hamilton C2 and C3 Intensive Care Ventilators
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Hamilton-C2 Intensive Care Ventilator |
Contact the manufacturer. |
160002 |
Hamilton-C3 |
Contact the manufacturer. |
160005 |
Issue
This field safety notice (FSN) provides an updated information on Hamilton medical ventilators Hamilton-C2/C3: Hamilton Medical AG became aware of a malfunction with our Hamilton-C2/C3 ventilators when used long term in neonatal patients. If an affected device Hamilton-C2 or Hamilton-C3 is used accumulatively for 91 days in neonatal patient without a restart of the device, the device will switch to "ambient state" and therefore stop ventilation. During the "ambient state" the device is alarming audibly until it is switched off by the user. Hence, it is mandatory to apply a periodic preventive restart of affected devices.
Devices and software version affected:
- All Hamilton-C2 with sw version ≤2.2.5 and neonatal option installed
- All Hamilton-C3 with sw version ≤2.0.9 and neonatal option installed
Recall start date: January 29, 2025