Health product recall

In2Flow Nasal Cannula, With OD15 Adapter

Last updated

Summary

Product
In2Flow Nasal Cannula, With OD15 Adapter
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

In2Flow Nasal Cannula, With OD15 Adapter

More than 10 numbers, contact manufacturer.

10076606
10076605
10076604

Issue

Hamilton Medical AG received several complaints where the plug (the one at the opposite side of the tubing) became detached from the In2Flow Nasal Cannula, causing leakage due to unintentional disconnection of that part. This may result in degraded high flow support and oxygen delivery, which could lead to a certain level of patient o2 desaturation.  

Affected product:  Hamilton Medical In2Flow Nasal Cannulas with a production date dated after 31.12.2022 and till January 2025 of all sizes and all lot numbers.

Recall start date: February 10, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Hamilton Medical Ag
Via Crusch 8, Bonaduz, Switzerland, 7402
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77025

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe