The ceiling tube support for the above device consists of an x-ray tube support pull-up section and an x-ray tube support section to which an x-ray tube assembly is mounted, and the tube holding section is secured to the column of the x-ray tube support…
RecallHealth product recall | 2023-05-19
Following Shimadzu's investigation, it was noted that there is a potential issue with the system software that could cause: 1) The x-ray irradiation rate (pulse rate) of pulse fluoroscopy to be different then set value, 2) The beam hardening filter…
RecallHealth product recall | 2022-04-12
As a result of Shimadzu investigation, it was determined that there is an issue with the adjustment procedure documented for the calibration of the system x-ray generator. This could cause x-ray radiation dose rate to exceed regulation and/or standards…
RecallHealth product recall | 2022-02-21