Health product recall

Radspeed Pro X-Ray and CH-200 Ceiling Tube Supports

Last updated

Summary

Product
Radspeed Pro X-Ray and CH-200 Ceiling Tube Supports
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Radspeed Pro X-Ray Tube Support

More than 10 numbers, contact manufacturer.

Not applicable.

CH-200 Ceiling Tube Support

More than 10 numbers, contact manufacturer.

Not applicable.

Issue

The ceiling tube support for the above device consists of an x-ray tube support pull-up section and an x-ray tube support section to which an x-ray tube assembly is mounted, and the tube holding section is secured to the column of the x-ray tube support pull-up section with eight bolts. It turned out that this bolt could come loose. If loosening of the bolt increases, the bolt may break. If all eight bolts break, the x-ray tube support section may eventually come off the column.

Recall start date: May 7, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Shimadzu Corporation

1, Nishinokyo-Kuwabaracho, Nakagyo-Ku, Kyoto-Fu, Japan, 604-8511

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73693

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe