Health product recall

Sonialvision G4

Last updated

Summary

Product
Sonialvision G4
Issue
Medical devices - Manufacturing defect
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products Lot or Serial Number Model or catalog number
Sonialvision G4 41C351653001, 41C3534B4001 503-78000-22, 503-78000-52

Issue

As a result of Shimadzu investigation, it was determined that there is an issue with the adjustment procedure documented for the calibration of the system x-ray generator. This could cause x-ray radiation dose rate to exceed regulation and/or standards.

Recall start date: Feb 9, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Shimadzu Corporation

1, Nishinokyo-Kuwabaracho, Nakagyo-Ku, Kyoto-Fu

Japan, 604-8511

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63939

Get notified

Receive notifications for new and updated recalls and alerts by category.