GE Healthcare has become aware that a diode on the power supply of the Prucka 3 Amplifier used with cardiolab/combolab systems could reach elevated temperatures. This could cause the amplifier to fail and consequently power off and become inoperable.…
RecallHealth product recall | 2023-08-21
Certain treadmills may accelerate forwards or backwards without notice. It may occur during initialization of the treadmill and within the first seconds of sending the start belt command. Also, the treadmill may come to a controlled yet unexpected stop.…
RecallHealth product recall | 2023-07-14
Full vision is conducting a corrective action related to the magnetic safety tether of the GE Healthcare T2100-ST Treadmill. When the magnetic safety tether is pulled, the treadmill may fail to stop resulting in a potential injury.
Recall start date:…
RecallHealth product recall | 2022-11-25
The Carescape Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station. Patient monitoring at the bedside is not affected.
Recall start date: May 10,…
RecallHealth product recall | 2022-05-16
Muse™ NX R1 systems with web edit option - edits can be lost after test is confirmed.
Recall start date: Mar 9, 2022
RecallHealth product recall | 2022-03-21
If the Carescape Central Station v2.0 is used with an unapproved keyboard that contains a mute key, and if the mute key is pressed, the audio will be muted resulting in loss of audible alarms. this issue can result in a delay in noticing a change in…
RecallHealth product recall | 2022-02-14