T2100 Treadmill
Last updated
Summary
Product
T2100 Treadmill
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Industry
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| T2100 Treadmill | Contact the manufacturer. | 2097357-002 2097357-001 |
Issue
Certain treadmills may accelerate forwards or backwards without notice. It may occur during initialization of the treadmill and within the first seconds of sending the start belt command. Also, the treadmill may come to a controlled yet unexpected stop. If the patient is standing on the belt, the patient may fall and suffer an injury.
Recall start date: May 24, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Physical medicine
Companies
GE Medical Systems Information Technologies Inc.
9900 Innovation Drive, Wauwatosa, Wisconsin, United States, 53226
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-73955