Displaying 1 - 5 of 5 items.
There is currently a non-conformance regarding the disinfection method recommended in the instructions manual for devices in the field, as the Keredusy disinfection method was not adequately validated. Long term effects of this method on the device…
RecallHealth product recall | 2023-07-10
Philips Respironics has detected environmental contamination (e.g. External dust and dirt) in the air path of some devices that have been returned from the field. Extended exposure to these environmental contaminants can lead to buildup of particulate on…
RecallHealth product recall | 2023-05-10
Philips Respironics has detected environmental contamination (e.g. external dust and dirt) in the air path of some devices that have been returned from the field. Extended exposure to these environmental contaminants can lead to buildup of particulate on…
RecallHealth product recall | 2023-04-26
Philips Respironics has discovered that accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FioO2 sensor may indicate a value…
RecallHealth product recall | 2023-03-29
The replacement silicone sound abatement foam installed into the Trilogy 100 and Trilogy 200 devices identified in this letter may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air…