Trilogy EVO and EV300
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Trilogy EVO, Canada |
All lots. |
CA2110X12B |
Trilogy EVO, O2, Canada EV300 |
All lots. |
CA2200X12B |
Issue
Philips Respironics has become aware that Trilogy EVO and Trilogy EV300 devices do not comply with the obstruction alarm requirements specified within ISO 80601-2-12 clause 201.12.4.108 and ISO 80601-2-72 clause 201.12.4.107.
The standards specify that the maximum delay from obstruction to alarm shall be no more than 2 breath cycles or 5 seconds, whichever is greater. The current obstruction alarm delay for these devices has been measured up to 65 seconds in certain bench test conditions. This is a delay potentially 60 seconds greater than required by the standards.
Recall start date: July 1, 2024
Additional information