Trilogy EVO and EV300
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Trilogy EVO, Canada |
All lots. |
CA2110X12B |
Trilogy EVO, O2, Canada EV300 |
All lots. |
CA2200X12B |
Issue
Philips Respironics has become aware that Trilogy EVO and Trilogy EV300 devices do not comply with the obstruction alarm requirements specified within ISO 80601-2-12 clause 201.12.4.108 and ISO 80601-2-72 clause 201.12.4.107.
The standards specify that the maximum delay from obstruction to alarm shall be no more than 2 breath cycles or 5 seconds, whichever is greater. The current obstruction alarm delay for these devices has been measured up to 65 seconds in certain bench test conditions. This is a delay potentially 60 seconds greater than required by the standards.
Recall start date: July 1, 2024
Additional information
Details
Respironics Inc. Also Trading As Philips Rs North America LLC
1001 Murry Ridge Lane, Murrysville, Pennsylvania, United States, 15668
Get notified
Receive notifications for new and updated recalls and alerts by category.