Health product recall

Philips Respironics Trilogy EV300 and Evo ventilator

Last updated

Summary

Product
Philips Respironics Trilogy EV300 and Evo ventilator
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalog number

Trilogy EV300

All lots.

CA2200X12B

Trilogy EVO

All lots.

CA2110X12B

Issue

Philips Respironics has detected environmental contamination (e.g. external dust and dirt) in the air path of some devices that have been returned from the field. Extended exposure to these environmental contaminants can lead to buildup of particulate on a flow sensor within the device. This may cause the device to inaccurately deliver pressure or volume/flow.

Recall start date: March 28, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Respironics Inc. Also Trading As Philips Rs North America LLC

1001 Murry Ridge Lane, Murrysville, Pennsylvania, United States, 15668

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-73573

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe