Philips Respironics Trilogy EV300 and Evo ventilator
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Trilogy EV300 |
All lots. |
CA2200X12B |
Trilogy EVO |
All lots. |
CA2110X12B |
Issue
Philips Respironics has detected environmental contamination (e.g. external dust and dirt) in the air path of some devices that have been returned from the field. Extended exposure to these environmental contaminants can lead to buildup of particulate on a flow sensor within the device. This may cause the device to inaccurately deliver pressure or volume/flow.
Recall start date: March 28, 2023
Additional information
Details
Respironics Inc. Also Trading As Philips Rs North America LLC
1001 Murry Ridge Lane, Murrysville, Pennsylvania, United States, 15668
Get notified