Displaying 1 - 14 of 14 items.
Issue 1 - Indication of block movements function:
When the operator activates the functions "block movements" or "block table movements" by pressing the block movements button on the pilot control module, there will be a dedicated message displayed on…
RecallHealth product recall | 2022-05-09
Shift of FMRI overlay and no direct possibility to correct the alignment contribute to an incorrect surgery planning when not detected:
- The original error is not generated in the software but a manual error during the scan.
- Incorrect anterior/…
RecallHealth product recall | 2022-05-03
During the transmission of the patient table position information, sporadic interference signals may occur due to an error within the patient table position encoder. This problem can be detected by the system and can result in a scan abort. Sporadic…
RecallHealth product recall | 2022-05-03
This update package provides the update VB40B_HF06 for Syngo.Via systems running on version VB40A or VB40B. This update solves several issues and introduces several improvements.
Recall start date: Apr 18, 2022
RecallHealth product recall | 2022-04-25
This update package provides the update VB30B for Syngo.Via systems running on version VB30A and any patch level.
Recall start date: Apr 14, 2022
RecallHealth product recall | 2022-04-25
Siemens Healthcare will be distributing new operator's manuals to affected customers which are needed for the upcoming software upgrade to VF11.
5 cautions were added to the manual.
Caution 1 (added two sentences to already existing caution): cause:…
RecallHealth product recall | 2022-04-11
The service pack (full version and delta package) NX VA20A-SP06 (VA20A-04T4) contains improvements concerning performance and stability.
- Image reconstruction failure solved
- Petra sequence image text corrected
- New LW for RCCS_2G to support new…
RecallHealth product recall | 2022-04-11
There is a potential issue with Artis systems in Syngo application software version VE20. After CT image data from toshiba is loaded, unintended image mirroring can occur along the horizontal and vertical image axes.
Recall start date: Mar 4, 2022
RecallHealth product recall | 2022-03-21
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the x-ray tube is no longer sufficiently cooled and the system will display the message "tube hot, have a break". Several minutes later the…
RecallHealth product recall | 2022-03-14
This is a performance update to system software VA10F and provides improvements of customer workflow and stability as well as remote update handling (RUH) functionality and cybersecurity updates.
Recall start date: Mar 8, 2022
RecallHealth product recall | 2022-03-14
Under certain sporadic circumstances the Sensis Vibe Hemo System with VD12A software the Co (cardiac output) measurement using the thermodilution method will temporarily no longer be possible. This happens without any obvious reason. When the issue…
RecallHealth product recall | 2022-01-31
Decreased head image quality on P46 systems after VB20 SP5 installation. This customer advisory notice provides a detailed error and workaround description.
Recall start date: Jan 6, 2022
RecallHealth product recall | 2022-01-17
An issue observed when creating a new GLM series with MOCO (motion correction) and /or spatial filtering using the MR Neuro 3D calculation in the MR Neuro3d workflow.
Recall start date: Jan 6, 2022
RecallHealth product recall | 2022-01-17
According to the Lungrads guideline the "increased" value should not appear in the table for rule 5 for non-solid lesions which have a mean total >= 20mm in the SWRS for all Lungrads finding types.
Recall start date: November 12, 2021…
RecallHealth product recall | 2022-01-04