Sensis Vibe
Brand(s)
Last updated
Summary
Product
Sensis Vibe
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
|---|---|---|
Sensis Vibe |
103218, 103221, 103266, 103268, 103220 |
11007641 |
Issue
Under certain sporadic circumstances the Sensis Vibe Hemo System with VD12A software the Co (cardiac output) measurement using the thermodilution method will temporarily no longer be possible. This happens without any obvious reason. When the issue appears on the Sensis Vibe Hemo system the syringe icon will no longer come up in the Co thermodilution dialog.
Recall start date: Nov 29, 2021
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Siemens Healthcare Gmbh
Henkestr. 127, Erlangen
Germany, 91052
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63873
Get notified