Artis Icono
Brand(s)
Last updated
Summary
Product
Artis Icono
Issue
Medical devices - Software defect
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
|---|---|---|
Artis Icono Biplane |
Not applicable. |
11327600 |
Artis Icono Floor |
Not applicable. |
11327700 |
Syngo Application Software |
Not applicable. |
11327667 |
Issue
There is a potential issue with Artis systems in Syngo application software version VE20. After CT image data from toshiba is loaded, unintended image mirroring can occur along the horizontal and vertical image axes.
Recall start date: Mar 4, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Health products - Medical devices - Radiology
Companies
Siemens Healthcare Gmbh
Henkestr. 127, Erlangen
Germany, 91052
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-64008
Get notified