Health product recall

Novum Iq Lvp

Last updated

Summary

Product
Novum Iq Lvp
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Novum Iq Lvp More than 10 numbers, contact manufacturer. 40700BAX

Issue

Baxter Corporation is issuing an urgent medical device correction for a subset of novum Iq large volume pumps (LVP) due to reports of pump operators pressing a single key on the novum keypad, but the pump registers the intended key as well as incorrectly registers an adjacent key.

Recall start date: August 8, 2023

Additional information

Previous recalls or alerts
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

 

Baxter Healthcare Corporation
1 Baxter Parkway, Deerfield, Illinois, United States, 60015
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-74111

Get notified

Receive notifications for new and updated recalls and alerts by category.