Novum IQ Large Volume Pump
Brand(s)
Last updated
Summary
Product
Novum IQ Large Volume Pump
Issue
Medical devices - Performance issue
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Novum IQ Large Volume Pump |
All lots. |
40700BAX |
Issue
Through ongoing internal testing, Baxter Corporation has observed isolated incident of over infusion on the Novum IQ LVP infusion pump during test runs at 0.1 ml/hr after 40 hours of continuous infusion within the 96-hour set change interval.
Recall start date: September 15, 2022
Additional information
Details
Original published date: 2022-09-23
Alert / recall type
Health product recall
Category
Health product - Medical device - General hospital and personal use
Companies
Baxter Healthcare Corporation
1 Baxter Parkway, Deerfield, Illinois, United States, 60015
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-64580