Dose IQ Safety Software
Brand(s)
Last updated
Summary
Product
Dose IQ Safety Software
Issue
Medical devices - Performance
What to do
Contact manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Dose IQ Safety Software |
All lots. |
DOSEIQW0001 |
Issue
Baxter Corporation is issuing an urgent medical device correction due to a defect in the Dose IQ Safety Software used with Novum IQ Large Volume Pump, resulting in an invalid initial setting for the air-in-line threshold for any new care area created in which none of the settings are changed.
Recall start date: May 29, 2023
Additional information
Previous recalls or alerts
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Baxter Healthcare Sa
Thurgauerstrasse 130, Glattpark (Opfikon), Switzerland, 8152
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-73759