Health product recall

Magec 2 Device

Last updated

Summary

Product
Magec 2 Device
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Magec 2 (Marketed As Magec X) Device (Various Rod Sizes And Lot Numbers)

More than 10 numbers, contact manufacturer.

MC2-5590S
MC2-6090S
MC2-4570S
MC2-5070S
MC2-4590S
MC2-5090S
MC2-5570S
MC2-6070S

Magec 2 (Marketed As Magec X) Device (Various Rod Sizes And Lot Numbers)

More than 10 numbers, contact manufacturer.

MS2-6090R
MC2-4590R
MC2-5070R
MC2-4570R
MC2-5590R
MC2-6070R
MC2-5570R
MC2-5090R

Issue

Recall relates to a packaging closure integrity issue. Investigation revealed a small population (3.8%) of packaged units have an undersized o-ring gland diameter that can affect the o-ring compression when the hard cap is affixed to the packaging tube. This lack of full compression can create a situation where the integrity of the container closure system is compromised potentially impacting the sterile barrier system.

Recall start date: November 1, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies

Nuvasive Specialized Orthopedics, Inc.

101 Enterprise, Aliso Viejo, California, United States, 92656

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72218

Get notified

Receive emails about new and updated recall and safety alerts.