Displaying 1 - 3 of 3 items.
Recall relates to a packaging closure integrity issue. Investigation revealed a small population (3.8%) of packaged units have an undersized o-ring gland diameter that can affect the o-ring compression when the hard cap is affixed to the packaging tube.…
RecallHealth product recall | 2022-11-24
Post-implantation fracture of an internal metallic component (i.e., locking pin) had been observed in early version of Magec rods that were manufactured prior to March 26, 2015. In June 2019, Nuvasive Specialized Orthopedics issued a safety alert to…
RecallHealth product recall | 2022-02-14
NSO is notifying that the products outlined are not indicated for use in individuals under the age of 18 years old. Also, NSO intends to make users aware of the gaps in the biological assessments for these devices as outlined in ISO 10993-1:2018…