Karl Storz Injection Needles
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Flexible Injection Needle |
All lots. |
27182A |
Injection Needle |
All lots. |
26175 PD |
Laryngeal Injection Needle |
All lots. |
8598B |
Issue
Validation testing of the efficacy of the manual cleaning for the injection needle was performed. The testing showed that the required efficacy level was not achieved. Different scenarios were tested, and all possible manual cleaning steps included, and the residual protein value remained above the acceptance criteria.
Recall start date: October 16, 2023
Additional information
Details
Karl Storz SE & Co. KG
Dr.-Karl-Storz Strasse 34, Tuttlingen, Germany, 78532