Validation testing of the efficacy of the manual cleaning for the injection needle was performed. The testing showed that the required efficacy level was not achieved. Different scenarios were tested, and all possible manual cleaning steps included…
RecallHealth product recall | 2023-10-31
Due to a failed revalidation of the sterilization process, there is no proof of the sterility of the products. The hazardous situation, that a contaminated product is used could lead to the consequence of patient infection.
There is no subsequent risk…
RecallHealth product recall | 2023-08-16
Instructions for use Z24357US, Revision BA inadvertently referred to the incorrect cleaning brush model in the Detergent Cleaning instructions. The affected instructions for use were posted online and accompanied products between April 18 and April 29,…
RecallHealth product recall | 2022-09-20
Karl Ttorz Endoscopy Canada is conducting a field action for specific serial numbers of C-View endoscopes with the potential to lose image when monopolar electrocautery is activated.
Recall start date: May 2, 2022
RecallHealth product recall | 2022-05-09
This recall notice describes a labelling update to correct the instructions for use for the flexible endoscopes identified to advise that high-level disinfection and, for specific models, liquid chemical sterilization, should not be used as methods of…
RecallHealth product recall | 2022-04-11