Health product recall

epoc Blood Analysis BGEM Test Card

Brand(s)
Last updated

Summary

Product
epoc Blood Analysis BGEM Test Card
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalogue number

Epoc Blood Analysis Bgem Test Card

Lot with prefix 02 or 12

10736515

Epoc Blood Analysis System - Single Use Blood Gas, Electrolyte And Metabolite Test Card

10736382, Lot prefix 00 or 06

CT-1006-00-00

Issue

Siemens confirmed that the occurrence of discrepant high pH results in samples introduced with higher injection volumes (beyond the epoc reader beep or after the "analyzing sample" message is displayed) with epoc sensor configuration 45.n (software version: epoc host v3.41.2 and epoc nxs v4.14.9 & 4.14.11)

Recall start date: March 25, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Epocal Inc.
2060 Walkley Road, Ottawa, Ontario, Canada, K1G 3P5
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-77251

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe