Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under specific conditions when the (automatic) delete blood tests feature is enabled. When the results are run on the epoc® NXS host and delete blood tests feature is…
RecallHealth product recall | 2023-11-17
Siemens Healthcare Diagnostics Inc. has confirmed the occurrence of discrepant high glucose results in samples with glucose results on the lower end of the reportable range (at or below 54 mg/dl) when using the above listed lots of BGEM test cards in…
RecallHealth product recall | 2023-06-15
Siemens Healthcare Diagnostics Inc. has confirmed the occurrence of inconsistent discrepant (low bias) po2 results. at a medical decision level of approximately 80mmhg, the maximum negative bias is - 11.7% (the acceptable bias is +/-10% for po2 range 50-…
RecallHealth product recall | 2022-06-20