Epoc Bgem Bun Test Card [25pk]
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot of serial number |
Model or catalogue number |
|---|---|---|
Epoc Bgem Bun Test Card [25pk] |
08-24092-60 |
10736515 |
Issue
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some card lots approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/l. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. Please note the card lots identified are now expired.
Recall start date: September 25, 2024
Additional information
Details
| Epocal Inc. |
| 2060 Walkley Road, Ottawa, Ontario, Canada, K1G 3P5 |