Health product recall

CareIntellect Perinatal

Last updated

Summary

Product
CareIntellect Perinatal
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

CareIntellect Perinatal

Version 2.3.1

K2504AA

Issue

GE Healthcare has become aware of an issue in CareIntellect Perinatal (CIP) that may result in loss of monitoring information and clearance of active patient monitoring alarms.

When this issue occurs due to an interruption in system connectivity, monitoring data will be unavailable for up to approximately 20 seconds and currently active visual and audible alarm indicators will no longer be displayed or announced at the CIP user interface. Following resolution of system connectivity, alarm triggers will restart based on the configured thresholds. This issue could delay recognition of a maternal or fetal condition requiring clinical assessment or intervention.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

GE Medical Systems Information Technologies Inc.

3114 N Grandview Blvd, Waukesha, Wisconsin, United States, 53188

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82711

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe