Product
Acl Top Cts
Issue
Medical devices - Inaccurate test or measurement results
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products Lot or Serial Number Model or catalog number
Acl Top 500 Cts All lots. 00000280040, 00000280040R
Acl Top 300 Cts All lots. 00000280060R, 00000280060
Acl Top 550 Cts All lots. 00000280045
Acl Top 350 Cts All lots. 00000280065

Issue

We have become aware of a potential carryover issue that may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when hemosil liquid anti-xa reagent is used in the same run with both hemosil liquid antithrombin and hemosil q.f.a. thrombin (bovine).

Recall start date: Feb 22, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies

Instrumentation Laboratory Co.

180 Hartwell Road, Bedford, Massachusetts

United States, 01730-2443

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63967