Search
Returning search results with filters:
Remove filter for
Brand: INSTRUMENTATION LABORATORY CO.
Clear all
Type
Audience
Category
Issue
Last updated
Recall class
Please note: using the search filters may not provide accurate search results for records prior to November 2021. When searching for older records, please ensure that you use the include archived checkbox.
Displaying 1 - 4 of 4 items.
Internal testing identified that the Gem Paks (cartridges) for the Gem Premier 3x00 with IGM, Gem Premier 4000 with IGM, and Gem Premier 5000 with IGM2 of impacted cartridge lots may experience an increase in CVP (Gem Premier 4000 Paks and Gem Premier…
RecallHealth product recall | 2023-11-06
An investigation into reported customer complaints identified that a sample misidentification could occur if the customer performs the following actions listed under the urgent medical device correction within a small window of time.
Recall start date…
RecallHealth product recall | 2023-08-15
Werfen is releasing a mandatory test library upgrade on the ACL Elite/Elite Pro (test library 02.05.00). This is a follow up to phase 1 of the recall for of the removal of Hemosil ® Liquid Anti-Xa (pn 0020302600 and 0020302601) on the ACL Elite/Elite Pro…
RecallHealth product recall | 2023-03-08
We have become aware of a potential carryover issue that may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when hemosil liquid anti-xa reagent is used in the same run with both hemosil liquid antithrombin and…
RecallHealth product recall | 2022-03-07