MINI-DAP, MICRO-DAP DIPHOTERINE (2019-04-19)
- Starting date:
- April 19, 2019
- Posting date:
- May 10, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69868
Last updated: 2019-05-10
Affected Products
- MINI-DAP DIPHOTERINE
- MICRO-DAP DIPHOTERINE
Reason
Damage inflected on the product (such as severe shocks), after delivery during the storage, lead to mechanical constraints between the valve and the pouch containing the solution which lead to a tear of the pouch that created a leak into the can. Consequently, the DIPHOTERINE solution was spread inside the canister and the MINI-DAP of DIPHOTERINE solution was not operating anymore.
Affected products
A. MINI-DAP DIPHOTERINE
Lot or serial number
Not applicable.
Model or catalog number
MINID
Companies
- Manufacturer
-
Prevor
Moulin de verville
Valmondois
95760
FRANCE
B. MICRO-DAP DIPHOTERINE
Lot or serial number
Not applicable.
Model or catalog number
MICROD