Health product recall

MINI-DAP, MICRO-DAP DIPHOTERINE (2019-04-19)

Starting date:
April 19, 2019
Posting date:
May 10, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69868


Last updated:

Affected Products

  1. MINI-DAP DIPHOTERINE
  2. MICRO-DAP DIPHOTERINE

Reason

Damage inflected on the product (such as severe shocks), after delivery during the storage, lead to mechanical constraints between the valve and the pouch containing the solution which lead to a tear of the pouch that created a leak into the can. Consequently, the DIPHOTERINE solution was spread inside the canister and the MINI-DAP of DIPHOTERINE solution was not operating anymore.

Affected products

A. MINI-DAP DIPHOTERINE

Lot or serial number

Not applicable.

Model or catalog number

MINID

Companies
Manufacturer

Prevor

Moulin de verville

Valmondois

95760

FRANCE



B. MICRO-DAP DIPHOTERINE

Lot or serial number

Not applicable.

Model or catalog number

MICROD