Acuson Juniper
Marque(s)
Résumé
Contactez le fabricant si vous avez besoin d'informations supplémentaires.
Produits visés
Produits touchés |
Numero de serie ou de lot |
Numero de modèle ou de catalogue |
---|---|---|
Acuson Juniper |
550822, 522941, 551868, 552635, 550751, 522943, 553951, 550562 |
11335791 |
Problème
Siemens Healthcare is notifying customer of a potential failure in the clip store function on the Acuson Juniper ultrasound system. In Juniper 1.0 the clip store function does not work when the ultrasound system has a disk full error. The purpose of this update is to provide juniper VA10G software to all juniper systems currently installed with software versions VA10D, VA10E and VA10F as a solution.
Date de rappel de debut: 22 janvier 2022