Acuson Juniper
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
---|---|---|
Acuson Juniper |
550822, 522941, 551868, 552635, 550751, 522943, 553951, 550562 |
11335791 |
Issue
Siemens Healthcare is notifying customer of a potential failure in the clip store function on the Acuson Juniper ultrasound system. In Juniper 1.0 the clip store function does not work when the ultrasound system has a disk full error. The purpose of this update is to provide juniper VA10G software to all juniper systems currently installed with software versions VA10D, VA10E and VA10F as a solution.
Recall start date: Jan 25, 2022