epoc® NXS Host – 5553CA
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
epoc® NXS Host – 5553CA |
More than 10 numbers, contact manufacturer. |
11413506 |
Issue
Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under specific conditions when the (automatic) delete blood tests feature is enabled. When the results are run on the epoc® NXS host and delete blood tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test.
Recall start date: October 24, 2023
Additional information
Details
Epocal Inc.
2060 Walkley Road, Ottawa, Ontario, Canada, K1G 3P5