Affected lot may contain the presence of particles
RecallHealth product recall | 2021-11-30
This voluntary medical device correction is being taken following the investigation of six customer reports stating that a total of four ventilators became inoperable during use. One of these reports indicated that, following the ventilator becoming…
RecallHealth product recall | 2021-11-29
Product sterility may be compromised in the affected lot due to the possibility of defects in the vial.
RecallHealth product recall | 2021-11-22
leica microsystems received one complaint (May 2021) pertaining to the M220 F12 surgical microscope system stating that the optics carrier fell down during the surgical procedure. The issue was found to be related to a loose screw in the…