Surgical Microscope Leica M220 F12
Brand(s)
Last updated
Summary
Product
Surgical Microscope Leica M220 F12
Issue
Medical devices - Mechanical defect
What to do
| Immediately stop using the product and contact manufacturer. |
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Surgical Microscope Leica M220 F12 |
70414002 |
10448367 |
Issue
|
leica microsystems received one complaint (May 2021) pertaining to the M220 F12 surgical microscope system stating that the optics carrier fell down during the surgical procedure. The issue was found to be related to a loose screw in the swing arm of the M220 F12 which is holding the optics carrier in place. Based upon this knowledge and a previous recall for the same issue this recall was opened.
Recall start date: 2021-10-21 |
What you should do
| Immediately stop using the product and contact manufacturer. |
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Companies
|
LEICA MICROSYSTEMS (SCHWEIZ) AG, MAX SCHMIDHEINY-STRASSE 201, HEERBRUGG, ST. GALLEN Switzerland |
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type I
Identification number
RA-63562
Get notified
Receive emails about new and updated recall and safety alerts.