In the identified lot number range of administration sets, the tubing outer diameter may vary in size. Additionally, the length of the tubing which contacts the air sensor may be too short resulting in the need to stretch the tubing to properly fit…
RecallHealth product recall | 2023-11-17
In the identified lot number range of administration sets, the tubing outer diameter may vary in size. Additionally, the length of the tubing which contacts the air sensor may be too short resulting in the need to stretch the tubing to properly fit…
RecallHealth product recall | 2023-11-07
Sub-assembly product code m43417505 was found to have the potential to leak due to poor concentricity and deformation in the male luer lock.
Recall start date: September 5, 2023
RecallHealth product recall | 2023-11-06
The extension set was noted during functional leak testing to have a high probability of potential failure which could lead to possible leakage or detachment at the device's bonded joint during use.
Recall start date: September 9, 2022
RecallHealth product recall | 2022-09-20
Investigations conducted at the manufacturing site identified that the lot may present an elevated risk of leakage from the air-inlet filter of the device.
Recall start date: Jan 25, 2022
RecallHealth product recall | 2022-02-14