Infusomat® Space Pump IV Set with Anti-Siphon Valve
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Infusomat® Space Pump IV Set With Anti-Siphon Valve |
lot numbers lower than 0061902124 |
363901 |
Infusomat® Space Pump Y-Type Blood IV Set With Anti-Siphon Valve |
lot numbers lower than 0061902124 |
363419 |
Infusomat® Space Pump Low Adsorption IV Set With Anti-Siphon Valve |
lot numbers lower than 0061902124 |
363902 |
Infusomat® Space Pump Burette IV Set With Anti-Siphon Valve |
lot numbers lower than 0061902124 |
363032 |
Issue
In the identified lot number range of administration sets, the tubing outer diameter may vary in size. Additionally, the length of the tubing which contacts the air sensor may be too short resulting in the need to stretch the tubing to properly fit within the Infusomat® space pump. If the tubing is stretched or the tubing outer diameter is too small, this may result in suboptimal coupling between the administration set and the air-in-line sensor dampening the signal of the air-in-line sensor. If the signal is dampened, the Infusomat® space pump may generate false air-in-line alarms.
Recall start date: October 23, 2023
Additional information
Details
B. Braun Medical Inc.
824 12th Avenue, Bethlehem, Pennsylvania, United States, 18018