Avanos medical is conducting a voluntary field correction for the Cortrak* 2 Enteral Access System (Eas) because modifications to the labeling of the device have been initiated. Reports of injuries and patient deaths related to misplacement of…
RecallHealth product recall | 2022-04-19
Smiths Medical became aware of eight (8) issues-(identified below) that can potentially occur that may impact therapy delivery in specific Medfusion 3500 and 4000 syringe infusion pumps.
1. Primary Audible Alarm (PAA)
2. Unanticipated depleted battery…
RecallHealth product recall | 2022-04-11
BBMI has identified through complaints the potential for fluid leakage or low fill volume of the respective containers of 0.9% sodium chloride injection USP 250ml [din : 01924303, lot no : l8002]. There is also the potential for smaller micro-leaks which…
RecallHealth product recall | 2022-03-14
The recall is being initiated due to a software defect that is triggered by disconnecting / connecting an USB device, causing a restart of the infinity central station and a temporary loss of central monitoring.
Recall start date: Mar 8, 2022
RecallHealth product recall | 2022-03-14