Infinity Centralstation Wide
Brand(s)
Last updated
Summary
Product
Infinity Centralstation Wide
Issue
Medical devices - Software defect
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
---|---|---|
Infinity Centralstation Wide |
More than 10 numbers, contact manufacturer. |
MS26800 |
Issue
The recall is being initiated due to a software defect that is triggered by disconnecting / connecting an USB device, causing a restart of the infinity central station and a temporary loss of central monitoring.
Recall start date: Mar 8, 2022
Additional information
Details
Original published date: 2022-03-17
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Draeger Medical Systems, Inc.
3135 Quarry Road, Telford, Pennsylvania
United States, 18969
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type I
Identification number
RA-63997