Health product recall

Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console (2021-03-05)

Starting date:
March 5, 2021
Posting date:
May 6, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75523

Last updated:

Produits Touchés

Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console

Reason

As a result of customer complaints, an issue was identified with a direct current to direct current (DC-DC) converter used within the system controller module of the Bio-Console 560 instrument. The issue has resulted in nine complaint occurrences related to issues with device performance including: the user interface not working, the instrument shutting down (no pump rotation), smoking and burning smell prior to and during use.

Affected products

Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console

Lot or serial number

560B102873
560B102874

Model or catalog number

560BCS1

Companies
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES