Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console (2021-03-05)
- Starting date:
- March 5, 2021
- Posting date:
- May 6, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75523
Last updated:
Produits Touchés
Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console
Reason
As a result of customer complaints, an issue was identified with a direct current to direct current (DC-DC) converter used within the system controller module of the Bio-Console 560 instrument. The issue has resulted in nine complaint occurrences related to issues with device performance including: the user interface not working, the instrument shutting down (no pump rotation), smoking and burning smell prior to and during use.
Affected products
Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console
Lot or serial number
560B102873
560B102874
Model or catalog number
560BCS1
Companies
- Manufacturer
-
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES