This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Zym B Reagent, API NH Strip, API Listeria, and FB Reagent (2014-03-07)

Starting date:
March 7, 2014
Posting date:
April 4, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38773

Recalled Products

A. Zym B Reagent

B. API NH Strip

C. API Listeria

D. FB Reagent

Reason

During the stability sudy of a new batch of Zym B reagent (70493), a visual defect has been observed. After three month of stability, this batch presents an heterogeneous aspect of color from yellow to red (a color yellow to amber is considered as conform according to the package insert). This visual defect is due to a specific new batch of raw material (lot number MKBP5202V of Fast Blue BB Salt Sigma) used for the manufacturing of several batches of FB reagent and Zym B reagent. As the involved batches of Zym B reagent are included inside of API NH (10400) and API Listeria (10300) kits, the both references are also impacted.

Affected products

A. Zym B Reagent

 

Lot or serial number

1002706740

1002717860

1002742610

1002760700

1002821420

1002857840

1002860530

 

Model or catalog number

70493

 

Companies
Manufacturer

Biomerieux Sa

Chemin De L'orme

Marcy-l'etoile

69280

FRANCE



B. API NH Strip

 

Lot or serial number

1002485910

1002485911

1002786740

1002809140

1002814890

1002814891

1002826750

1002869720

1002869721

 

Model or catalog number

10400

 

Companies
Manufacturer

Biomerieux Sa

Chemin De L'orme

Marcy-l'etoile

69280

FRANCE



C. API Listeria

 

Lot or serial number

1002569040

1002569041

1002587190

1002856790

1002856800

 

Model or catalog number

10300

 

Companies
Manufacturer

Biomerieux Sa

Chemin De L'orme

Marcy-l'etoile

69280

FRANCE



D. FB Reagent

Lot or serial number

1002717180

1002737220

1002778550

1002782960

1002792840

1002834130

1002877370

Model or catalog number

70562

Companies
Manufacturer

Biomerieux Sa

Chemin De L'orme

Marcy-l'etoile

69280

FRANCE